Recommendations and reminders

OUR PROBES ARE INDIVIDUALLY PACKAGED TO PROTECT THE PROBES UNTIL THEY WILL BE IN USE.

We recommend:

Avoid fingerprints or any organic matter on the surface of the probe itself to maintain the integrity of the probe. Wear gloves is a good practice.

There should be no humidity inside the probes before use.

Avoid breakage by enough preheating the probe before initialization of the immersing procedure in melted aluminum.

CAUTION

The temperature of melted aluminum is higher than the annealing temperature of the aluminosilicate glass. Meaning that a probe should be thermally stable until being removed from melted aluminum.

After having removed a probe from melted aluminum, remove the probe from the LiMCA's head because the probe will begin to crack, and contamination with dropping parts of aluminosilicate glass on the melted aluminum may occur.